| Facility | Introduction | Maintenance cycle | Operation and maintenance plan and schedule |
|---|---|---|---|
| Emergency shower and eyewash station | Scope: This refers to the testing of integrated emergency shower and eyewash stations (non-desktop type) within the completed laboratory area. Testing Items: eyewash station water pressure, regular water discharge from the shower, and annual shower water volume testing. | Biweekly | Click the link to view |
| Carbon dioxide fire extinguisher | Scope: This refers to the testing of carbon dioxide fire extinguishers supplied by the laboratory service department, including visual inspection and periodic weighing. | Monthly | Click the link to view |
| Fume Hood | Scope: This refers to the testing of completed and in-use laboratory fume hoods. Testing Items: face velocity, airflow visualization, and terminal function checks. | Yearly | Click the link to view |
| Biosafety cabinet | Scope: This refers to the testing of completed and in-use laboratory biosafety cabinets. Testing Items: airflow rate, HEPA performance, ultraviolet radiation, illuminance, airflow pattern, and noise level. | Yearly | Click the link to view |
| High-efficiency filters and FFU units | Scope: This refers to the testing of HEPA filters in completed and operational laboratory cleanrooms. Testing Items: HEPA filter replacement, FFU unit replacement, etc. | 2-4 years (depending on usage) | Click the link to view |